PROGRAMMING AND OPERATING THE IMY
PROGRAMMING THE IMY
Although the unit has been programmed at the factory, the set-ups will
generally have to be changed to suit the application. Basic set-up is complete
after selection of the temperature units, decimal point placement, and
selection of the digital filtering level.
Before actually trying to program the indicator, it is advised to organize all
the data for the programming steps to avoid any possible confusion and to
read the programming procedure entirely before proceeding.
To set-up the indicator, connect primary power and signal wires as
outlined in the connections section (Appendix “A”). Remove the jumper
wire (if installed) from TBA #3 (PGM. DIS.). This will allow the operator to
enter and modify all of the indicator’s parameters. Press the front panel
button labeled “P”, momentarily. Briefly, the display will show “Pro”
alternately flashing with “0”. This is the indicator’s programming mode.
The programming mode is divided into sections, numbered 0-9, each of
which can be individually accessed. The front panel “UP” and “DOWN”
arrow buttons can be used to select one of these numbers and the “P” button
can be used to enter the selected programming module. In all of the
programming modules, “UP” and “DOWN” are used to either select from a
list of choices or enter a value. The “P” button is used to save the new value
and progress to the next step within a module (Note: the new value takes
effect when “P” is pressed). Upon completion of a module, the indicator
returns to the “Pro” <>“0” stage. Pressing the “P” button at this time
causes the unit to display “End” after which the unit returns to the normal
display mode. The following table explains the basic function of each step.
Note: < > This indicates that the display will alternate between the English
prompt and the actual data.