PROGRAM FUNCTION OF E1-CON AND OPTIONAL E2-CON
(Cont’d)
“9” - If the alarm option is installed, a low level resets a latched or
unlatched alarm into its inactive state. This provides manual
override of alarms for system start-up and other unusual
events such as system testing.
“10” - A negative going edge toggles the display between “input”
and “total” (from input to total, or vice versa). No action is
taken on the positive going edge.
“11” - A negative going edge zeros (tares) the input signal and adds
the value that was in the input display to the totalizer value,
every time this operation is performed. The time-base, scale
factor and low cut-out in “Module #5” are in affect disabled,
when this function is selected.
“12” - Display hold with offset. A negative going edge tares (zeros)
the input signal. Prior to the offset operation, the input
signal is saved and held (display hold) as long as the remote
input pin is low. On the positive edge, the input display will
show zero. If there is an increase to the input signal while the
remote input is low, the display will reflect (show) the
increase at the positive edge.
“13”- Instrument reading synchronization. A low level disables all
meter operations (alarms, total, analog out, etc.). A positive
edge resets the start of the A/D conversion, to allow
synchronization with external processes and controls.
While in this function, the other E-CON pin will be
operational.
“14”- Print request. Transmits data according to the print options
that have been selected in Program Module #7. If the low
time exceeds 800 msec, a second print-out may occur.
“E2-CON”<>Ifthetotalizer option is installed, E2-CON has the
same programmable functions as E1-CON.